This is written for AccuVax T-1000 (Version 1.4.0)

Introduction and Overview

There are a number of reasons why an AccuVax needs to be power cycled.  This article will detail how to power off and power on AccuVax.


Powering off AccuVax

  1. Power off Tablet
  • Press Power button
    • If there's not a button, use a paperclip into the hole in the bezel at the top of the touchscreen.
  • Tablet will prompt user to "Slide to Power off" - user should use their finger to slide down
Tablet Power Location


  • Disengage Battery and Turn off Power
  • Pull the Battery Circuit Breaker out (note the picture below)
  • Flip power button to off







There should not be any noise coming from the unit.


Powering on AccuVax

  1. Engage Battery and Turn on Power
  • Push the Battery Circuit Breaker in (note the picture below)
  • Flip power button to on



  • Ensure tablet turns on (BIOS is set to wake on AC)
    • If the tablet doesn't power on, use the power button to turn it on and make a ticket to fix the BIOS settings.